YSI 5200 User Manual

Page 72

Advertising
background image

YSI Incorporated

70

5200 Recirculating System Monitor

12. Press the Enter [



] key to enter the value. The display will notify you that this is Pass(2), then

the pH Entry screen is displayed.

P a s s 2

p H = 7 . 0 4 T = 7 3 . 5 °
W a i t t i l S t a b l e

13. Repeat steps 4 through 10 above (pass 2) using a second pH buffer.

14. Press the Enter [



] key to enter the value. After displaying a brief message, the display will

return to the pH Calibration menu.

C a l i b r a t i o n

C o m p l e t e

C a l i b r a t i o n

T w o - P o i n t

15. Press the [

ESC

] key to exit the pH Calibration menu. You will be prompted to save your changes.

C a l i b r a t i o n
S a v e ? Y e s

16. Press the Enter [



] key to save the new calibration value, or press the [e] or [f] key to select

No, then press the Enter [



] key to exit without saving any changes.

17. Rinse the probe module, calibration/storage cup and sensors in tap or purified water and dry.