5200 Recirculating System Monitor

Figure 4-3 Port Plug Removal

3. Locate the port with the connector that corresponds to the sensor that is to be installed. The

number of pins is specific for each probe type.

Figure 4-4 Sensor Port Identification

4. Apply a thin coat of o-ring lubricant (supplied in the YSI 5511 maintenance kit) to the o-rings on

the connector side of the sensor (see Figure 4-5 O-Ring Lubrication).

O-Rings

Sensor nut

Figure 4-5 O-Ring Lubrication

CAUTION: Make sure that there are NO contaminants between the O-ring and the sensor.

Contaminants that are present under the O-ring may cause the O-ring to leak.

5. Be sure the probe module sensor port is free of moisture and then insert the sensor into the correct

port. Gently rotate the sensor until the two connectors align.

6. With connectors aligned, screw down the sensor nut by hand, and then tighten using the sensor

installation tool.

Figure 4-6 Sensor Installation

pH or pH/ORP port

Conductivity/Temperature
port

Oxygen port