Calibration procedures, Accessing the calibrate screen, Alibration – YSI 5200 User Manual

Page 64: Rocedures, 2 calibration procedures

Advertising
background image

YSI Incorporated

62

5200 Recirculating System Monitor

6.2 Calibration

Procedures

6.2.1

Accessing the Calibrate Screen

1. From the Run screen, press the Enter [



] key, then the [e] or [f] key until the Parameters

option appears.

P a r a m e t e r s

E n t e r t o S e l e c t

2. Press the Enter [



] key to select Parameters.

D O S y s t e m

E n t e r t o S e l e c t

3. Press the [e] or [f] key until the parameter you want to calibrate appears. Conductivity is used

as an example below.

C o n d S y s t e m

E n t e r t o S e l e c t

4. Press the Enter [



] key to display the Enabled screen.

C o n d S y s t e m

E N A B L E D ? N o

5. If the parameter is disabled (as indicated by No appearing on the display), press the [e] or [f]

key to select Yes.

C o n d S y s t e m

E N A B L E D ? Y e s

6. Press the Enter [



] key.

C a l i b r a t i o n

E n t e r t o S e l e c t

7. Press the Enter [



] key to select Calibration.

8. Follow the instructions in the following sections for the specific parameter that you are

calibrating.