5200 Recirculating System Monitor

Range High

From the Set Point menu (with the Range High option displayed), press the Enter [



] key.

R a n g e H i g h
0 7 8 . 8 ° F

Using the [e], [f] and Enter [



] keys, enter the Range High value.

NOTE: The Range High value must be higher than the Set Point.

R a n g e H i g h
0 7 4 . 0 ° F

Press the Enter [



] key. The display returns to the Set Point Menu.

E n t e r t o S e l e c t

R a n g e H i g h

Press the [f] key for the next Set Point selection, Alarm Low.

Alarm Low

From the Set Point menu (with the Alarm Low option displayed), press the Enter [



] key.

A l a r m L o w
0 6 8 . 8 ° F

Using the [e], [f] and Enter [



] keys, enter the Alarm Low value.

NOTE: The Alarm Low value must be lower than the Range Low.

A l a r m L o w
0 6 2 . 0 ° F

Press the Enter [



] key. The display returns to the Set Point Menu.

E n t e r t o S e l e c t

A l a r m L o w

Press the [f] key for the next Set Point selection, Alarm High.

Alarm High

From the Set Point menu (with the Alarm High option displayed), press the Enter [



] key.

A l a r m H i g h
0 7 9 . 7 ° F

Using the [e], [f] and Enter [



] keys, enter the Alarm High value.

NOTE: The Alarm High value must be higher than the Range High.

A l a r m H i g h
0 8 5 . 0 ° F

Press the Enter [



] key. The display returns to the Temp System Menu.