Alarm System

The alarm system for each measurement parameter can be turned on or off. When on, an audible alarm
will sound on the YSI 5200 and an alarm page will be sent (if a modem is installed and paging is
enabled). The alarm will remain on until a key is pressed. See 7.1 Alarm Logic Flow Charts for details.

Setup Menu

DO Setup Menu selections are shown below.

Setup

DIsplay Range

Display Format

Interval

Manual

Auto

Accept

Off, 10, 20, 30, 40, 50, 60

%Sat

MG/L

Accept

Accept

Low

High

Membrane Type

Teflon

PE

Accept

Accept

Figure 5-7 DO Setup Menu

Interval

The 5200’s DO system operates in either continuous or interval mode. In continuous mode, the DO probe
is on all the time. In interval mode, the DO system is powered down for a selected interval (10, 20, 30, 40,
50 or 60 minutes), then powered back up for 75 seconds. The DO level is measured during the last 15