Ph calibration, 4 ph calibration – YSI 5200 User Manual

Page 70

Advertising
background image

YSI Incorporated

68

5200 Recirculating System Monitor

6.2.4 pH

Calibration

Calibration

1 Point

Display Actual Value " " Current Temp Value

Scrolling Text*

2 Point

Display Actual Value " " Current Temp Value

Scrolling Text*

Display Actual Value " " Current Temp Value

Scrolling Text*

Calibration

Complete

Get UserValue

Figure 6-4 pH Calibration Menu

1. Go to the pH calibrate screen as described in Section 6.2.1 Accessing the Calibrate Screen.

E n t e r t o S e l e c t

1 - P o i n t

2. Select either One-Point or Two-Point calibration.

E n t e r t o S e l e c t

2 - P o i n t

Ö

Select the 1-point option only if you are adjusting a previous calibration. If a 2-point
calibration has been performed previously, you can adjust the calibration by carrying out a

Advertising