Ph calibration, 4 ph calibration – YSI 5200 User Manual
Page 70
Advertising

YSI Incorporated
68
5200 Recirculating System Monitor
6.2.4 pH
Calibration
Calibration
1 Point
Display Actual Value " " Current Temp Value
Scrolling Text*
2 Point
Display Actual Value " " Current Temp Value
Scrolling Text*
Display Actual Value " " Current Temp Value
Scrolling Text*
Calibration
Complete
Get UserValue
Figure 6-4 pH Calibration Menu
1. Go to the pH calibrate screen as described in Section 6.2.1 Accessing the Calibrate Screen.
E n t e r t o S e l e c t
1 - P o i n t
2. Select either One-Point or Two-Point calibration.
E n t e r t o S e l e c t
2 - P o i n t
Ö
Select the 1-point option only if you are adjusting a previous calibration. If a 2-point
calibration has been performed previously, you can adjust the calibration by carrying out a