Auxiliary inputs wiring, Example (digital)—installing a float switch, Uxiliary – YSI 5200 User Manual

Page 81: Nputs, Iring, 2 auxiliary inputs wiring, 1 example (digital)—installing a float switch

Advertising
background image

YSI Incorporated

79

5200 Recirculating System Monitor

Wire

Terminal

Connection

Solid Blue Wire

1

T (-)

White w/Blue Stripe

2

T (+)

Solid Orange Wire

3

R (-)

White w/Orange Stripe

4

R (+)


NOTE:
On the final slave in a string, install one 120

Ω (ohm) resistor between T(+) and T(-) and one

120

Ω (ohm) resistor between R(+) and R(-). Resistors are included with each 5200.

8.2 Auxiliary

Inputs

Wiring

The auxiliary inputs are configurable for a number of applications including monitoring level and/or
pressure switches. Both inputs have input protection, but the maximum input voltage should be no more
than 6 VDC. YSI recommends that twisted pair wire be used for these connections.

CAUTION: The auxiliary inputs are not isolated. Devices connected must be electrically isolated from

ground and the water.

CAUTION: Observe correct polarity on analog input. Do not apply negative voltage to analog input.

Two snap-on ferrite assemblies, YSI# 655361, are included with the 5200 Monitor. Wires being
connected to the Auxiliary ports need to be routed through the ferrites as shown in Figure 3-9. Failure to
do so could cause catastrophic monitor failure.

Auxiliary inputs can be configured as follows:

Input #1 – Digital (contact open/close w/current source)

Caution: Do not input external voltage.

Input #2 – User Defined (0 to 1.0 VDC, 0 to 5.0 VDC, or 4 to 20 mA selectable; max. 6.0 VDC) 10K

impedance

or

Digital (contact open/close w/current source)

Caution: Do not input external voltage.

NOTE: For proper digital input operation, use low impedance switches and wiring. The total resistance

should not exceed 200 ohms.

8.2.1

Example (Digital)—Installing a Float Switch

1. Install the float switch at the desired water level according to the manufacturer’s instructions.

T(-) T(+) R(-) R(+)

T(-) T(+) R(-) R(+)

Master

\

\ \ \

T(-) T(+) R(-) R(+)

Final slave

RS485 Wiring Diagram for 3 Networked 5200s

120

\

\

\

\

120

120

120

T(-) T(+) R(-) R(+)

All other slaves

\

\

\

\