General alarm, Display address (serial number), Data – YSI 5200 User Manual

Page 93: Data logger, Clear data, Print data, Entire log, Current log, 6 general alarm, 7 display address (serial number)

Advertising
background image

YSI Incorporated

91

5200 Recirculating System Monitor

12. Disconnect power from the 5200.

13. Reapply power to the 5200.

9.1.6 General

Alarm

See Section 7.2 General Alarm for a detailed description.

9.1.7

Display Address (Serial Number)

Displays the Address (serial number) of the 5200 on the screen. The Address and software version are
also displayed whenever the 5200 is powered on.

9.1.8 Data

The 5200 has a 32K volatile memory, allowing it to log 750 sample data sets. The logging rate is user
selectable. Data menu selections are Data Logger, Clear Data, Print Data and Log Events.

Data Logger

Select Data Logger to enter the desired Data Logging time interval.

Clear Data

Clear Data erases all logged data stored in the 5200’s memory.

Print Data

Use Print Data to send logged sample data via the 5200’s RS232 serial port to a computer equipped with
any standard communications software. See Section 3.7.1 RS232 and TCP/IP Port for wiring
connections.

Entire Log

Sends the Entire Log to a computer via the 5200’s RS232 port. The output is tab delimited ASCII text
that can be imported into a spreadsheet. The format is shown below. Headings are not included in the
output.

Date

Time

pH

ORP

Temp

Cond

DO

DO% A1A2 AC CS


05/17/2001

09:32:29AM

6.97

209mV

76.6F

187uS

17.22 207

0 0 0 136

05/17/2001

09:38:04AM

6.97

209mV

76.8F

221uS

13.04 157

0 0 0 146

05/17/2001

09:43:04AM

6.97

209mV

76.8F

227uS

12.80 154

0 0 0 144

05/17/2001

09:48:04AM

6.98

209mV

76.8F

249uS

12.70 153

0 0 0 136

A1—Auxiliary Input #1. 0 or 1
A2—Auxiliary Input #2. Digital mode, 0 or 1
AC— Auxiliary Input #2. Analog mode, Analog Counts (4096 counts = full scale)
CS—Check Sum

Current Log

Sends the Current Log (current readings) to a computer via the 5200’s RS232 port. The format is
identical to the Entire Log format shown above. Note that the data sent is the current reading from the
sensors not the last logged data point.