2.4 Instrument Mounting


The 4014B and 4040B Function Generators are intended
for bench use. The instrument includes a front feet tilt
mechanism for optimum panel viewing angle. The
instrument does not require special cooling when
operated within conventional temperature limits. It may
be installed in a closed rack or test station if proper air
flow can assure removing about 15 W of power
dissipation.

2.5 Power Requirements

The function generators can be operated from any
source of 90V to 264V AC, frequency from 48Hz to 66Hz.
The maximum power consumption is 25
VA(4014B)/30VA(4040B). Use a slow blow fuse of 1A,
UL/CSA approved as indicated on the rear panel of the
instrument.

The instrument power fuse is located in the AC input

plug. To access the fuse, first disconnect the power cord
and then remove the fuse box.