tells the device to return the minimum or maximum value
to which the parameter may currently be set. For example,

FREQ? MAX


will return the maximum value to which the

frequency may currently be set.

Not all Program Message units have query forms (for
example, SAV), and some Program Message Units might
have only the query form (for example IDN?).

The instrument puts the response to the query into the
output queue, from where it may be read by the controller.

NOTE: All commands should be terminated with a
<CR> carriage return or <LF> line feed character. Any
response string from a query command has both
<CR> and <LF> characters appended. For example,
an amplitude query command will return
1.23<CR><LF>.

4.7 Status Reporting


The instrument is capable of reporting status events and
errors to the controller.

4.7.1 The Error Queue


The error queue is used to store codes of errors detected
in the device. It is implemented as a cyclic buffer of

Advertising