8 common commands, 1 system data commands – B&K Precision 4040B - Manual User Manual

Page 61

Advertising
background image


61

Data were sent before the entire

response of a previous query was read.

-420

Query UNTERMINATED.

An attempt was made to read a response before the
complete program message meant to generate that
response was sent.

Warnings


The execution of some commands might cause an
undesirable instrument state. The commands are
executed,

but a warning is issued.


500

Trigger rate short

510

Output overload


"Trigger rate short" means that the period of the waveform
is larger than the value of the internal trigger rate. Thus
not every trigger will generate a cycle (or burst) of the
waveform.

4.8 COMMON COMMANDS

4.8.1 System Data Commands


*IDN? - Identification query

The identification query enables unique identification of
the device over the GPIB. This query should always be
the last in a program message. It returns a string with four
fields: