1) The MAXimum amplitude is dependent on the offset.
2) The MAX and MIN arguments should not be used in a
program message containing an OFFSet command, since
these values are evaluated during parsing, based on the
current value of the offset.


OFFSet <offset>

The offset command is used to set the DC offset of the
output waveform. Note that the amplitude and the offset
are limited by the relation:

Peak Amplitude + |Offset| ≤ 5V


Arguments
Type:

Numeric

Units:

V, mV

Range:

10mV to 5V

Rounding: To 10mV
Command Type: Setting or Query
Setting
Syntax:

OFFSet<ws><offset>[units]

OFFSet<ws>MINimum|MAXimum
Examples:

OFFS 2.5

OFFS 2.5V
OFFS MAX
Query
Syntax:

OFFSet?[<ws>MINimum|MAXimum]

Examples:

OFFS?

OFFS? MAX