3 am shape, 4 am frequency, 5 am source – B&K Precision 4078 - Manual User Manual

Page 60

Advertising
background image

Rounding:

to integer

Command Type: Setting or Query
Setting
Syntax:

:SOURce:AM:DEPTh<ws><percent depth>

:SOURce:AM:DEPTh<ws>MINimum|MAXimum

Examples:

AM:DEPTh 50

Query
Syntax:

AM:DEPTh?[<ws>MINimum|MAXimum]

Response:

NR3

4.13.1.7.3 AM SHAPe

This command selects the AM modulating waveform shape

Arguments
Type:

Character

Options:

SINusoid, TRIangle, SQUare

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]AM:SHAPe<ws><SIN|TRI|SQU>

Examples:

[:SOURce:]AM:SHPE SIN

AM:SHAPE TRI

Query
Syntax:

[:SOURce:]AM:SHAPe?

Response:

SIN|TRI|SQU

4.13.1.7.4 AM FREQuency


This command sets the AM modulating waveform frequency

Arguments
Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Fmax = 20 KHz

Fmin = 0.01 Hz

Rounding:

The value is rounded to 4 digits.

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]AM:FREQuency<ws><frequency>[units]

[:SOURce:]AM:FREQuency<ws>MINimum|MAXimum

Examples:

AM:FREQ 5KHZ

AM:FREQ 5E3

AM:FREQ MAXIMUM

AM:FREQ MIN


Query
Syntax:

[:SOURce:]AM:FREQuency?[<ws>MAXimum|MINimum]

Examples:

AM:FREQ?

AM:FREQ? MAX

Response:

NR3

4.13.1.7.5 AM SOURce


This command selects the AM modulation source as either internal (then the above settings are effective) or external (and
then the external waveform determines depth, shape and frequency of modulation).

Arguments
Type:

Character

Options:

INTernal, EXTernal

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:] AM:SOURce<ws><INT|EXT>