Examples:

AM:SOUR INT

AM:SOUR EXT

Query
Syntax:

[:SOURce]:AM:SOURce?

Response:

INT|EXT

4.13.1.8 FM modulation

The following commands control the FM modulation:

4.13.1.8.1 FM STATe

This command activates or deactivates FM modulation:

Arguments
Type:

Boolean

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]FM[:STATe]<ws>ON|1|OFF|0

Examples:

FM:STAT ON

FM OFF

Query
Syntax:

[:SOURce:]FM[:STATe]?

Response:

0|1

4.13.1.8.2 FM DEViation

This command sets the FM modulation deviation

Arguments
Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Dependent on the carrier frequency.

Fmax = carrier frequency

Fmin = 10 uHz

Rounding:

The value is rounded to 4 digits.

Command Type: Setting or Query
Setting
Syntax:

[:SOURce]:FM:DEViation<ws><frequency>[units]

[:SOURce]:FM:DEViation<ws>MINimum|MAXimum

Examples:

FM:DEV 5KHZ

FM:DEV 5E3

FM:DEV MAXIMUM

FM:DEV MIN

Query
Syntax:

[:SOURce]:FM:DEViation?[<ws>MAXimum|MINimum]

Examples:

FM:DEV?

FM:DEV? MAX

Response:

NR3

4.13.1.8.3 FM SHAPe

This command selects the FM modulating waveform shape

Arguments
Type:

Character

Options:

SINusoid, TRIangle, SQUare

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]FM:SHAPe<ws><SIN|TRI|SQU>

Examples:

[:SOURce:]FM:SHPE SIN

FM:SHAPE TRI

Query