B&K Precision 4078 - Manual User Manual

Page 96

Advertising
background image


90

Amplitude

setting

Minimum reading

RMS

Maximum reading RMS DVM reading RMS

10Vp-p

3.499V

3.572V

5Vp-p

1.749V

1.786V

3Vp-p

1.049V

1.072V

1Vp-p

0.349V

0.358V

100mVp-p

34 mV

37 mV

50mVp-p

17 mV

19 mV

5. Set the unit to generate a SQUARE wave with 1KHz frequency.

6. CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and

50mVp-p are in the accuracy range:

Amplitude

setting

Minimum reading

RMS

Maximum reading RMS DVM reading RMS

10Vp-p

4.949V

5.051V

5Vp-p

2.474V

2.526V

3Vp-p

1.484V

1.516V

1Vp-p

0.494V

0.51

100mVp-p

49 mV

52 mV

50mVp-p

24 mV

26 mV

4

Set the unit to the ARBITRARY function and generate a sine wave signal:

- Press the ARB key

- Select F2 – ARB

- Select F4 – EDIT

- Select F3 – PREDEF

- Select F2 – FROM, press key 1 and ENTER

- Select F2 – DATA, press key 0 and ENTER

- Select F3 – LENG, press 1000 and ENTER

- Select F3 – SCALE, press 100 and ENTER

- Select F4 – EXEC

- Select F3 – YES

- Select ARB

- Select F1 – FREQ, press 1000 Hz