:SYSTem:ERRor?

This query returns the first entry in the error queue, and removes that entry from the queue. Its function is identical
to that of the :STATus:QUEue:NEXT? query.

Command Type: Query only
Query
Syntax:

:SYSTem:ERRor?

Response:

<Error number>, "<error description>"


4.13.6.3 SCPI Version


:SYSTem:VERSion?


This query is used to read the SCPI version to which the instrument complies.

Command Type: Query only
Query
Syntax;

:SYSTem:VERSion?

Response:

1992.0 (NR2 format)

4.13.6.4 Security


:SYSTem:SECurity[:STATe] <Boolean>

This command enables the instrument memory to be cleared. The stored settings and the arbitrary waveform
memory are cleared when the Security state is changed from ON to OFF. The instrument state is returned to the
factory power-on default.

Arguments
Type:

Boolean

Command Type: Setting or Query
Setting
Syntax:

:SYSTem:SECurity[:STATe]<ws>ON|1|OFF|0

Examples:

:SYST:SEC ON

:SYST:SEC OFF

Query
Syntax:

:SYSTem:SECurity[:STATe]?

Response:

0|1

4.13.6.5 Power-on Buffer


:SYSTem:POBuffer <buffer number>

This command is used to set the Power On Buffer setting. The instrument will power-on with the setting stored in
that buffer.

Arguments
Type:

Numeric

Range:

0 to 49

Rounding:

to integer value

Command Type: Setting or Query
Setting
Syntax:

:SYSTem:POBuffer<ws><buffer>|MINimum|MAXimum

Example:

:SYST:POB 99

Query