10 start address, 11 wavelength, 12 marker address – B&K Precision 4078 - Manual User Manual

Page 76

Advertising
background image

4.13.4.10 Start Address


:ARBitrary:STARt <start address>


This command sets the start address of the waveform to be run.

Arguments
Type:

Numeric

Range:

1 to 399,999

Rounding:

to integer value

Command Type: Setting or Query
Setting
Syntax:

:ARBitrary:STARt<ws><start address>

:ARBitrary:STARt<ws>MINimum|MAXimum

Example:

:ARB:STAR 100

Query
Syntax:

:ARBitrary:STARt?[<ws>MINimum|MAXimum]

Examples:

:ARB:START?

:ARB:STAR? MIN

Response:

NR1

Considerations: The start address and length must meet the condition:


Start Address + Length - 1

≤ 400,000

4.13.4.11 Wavelength


:ARBitrary:LENGth <length>


This command sets the length of the waveform being run.

Arguments
Type:

Numeric

Range:

2 to 400,000

Rounding:

to integer value

Command Type: Setting or Query
Setting
Syntax:

:ARBitrary:LENGth<ws><length>

:ARBitrary:LENGth<ws>MINimum|MAXimum

Example:

:ARB:LENG 1E3

Query
Syntax:

:ARBitrary:LENGth?[<ws>MINimum|MAXimum]

Example:

:ARB:LENG?

Response: NR1
Considerations: 1) Changing the wavelength will change either the frequency.

2) The minimum wavelength is 2.

4.13.4.12 Marker Address


:ARBitrary:MARKer [:ADDRess] <marker address>


This command is used to set the address of the marker.

Arguments
Type:

Numeric.

Range:

1 to 400,000

Rounding:

to integer values