3 sweep time, 4 sweep start, 5 sweep stop – B&K Precision 4078 - Manual User Manual

Page 65

Advertising
background image

4.13.1.10.3 Sweep TIME


This command sets the time for one complete sweep:

Arguments
Type:

Numeric

Units:

S, mS, uS, nS

Range:

20mS to 500S

Rounding:

to 4 digits

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]SWEEP:TIME<ws><time>[units]

[:SOURce:]SWEEP:TIME<ws>MINimum|MAXimum

Examples:

SWEEP:TIME 50MS

Query
Syntax:

[:SOURce:]SWEEP:TIME?[<ws>MINimum|MAXimum]

Response:

NR3


4.13.1.10.4 Sweep STARt


This command sets the start frequency of the sweep:

Arguments
Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Dependent on the frequency range of the current function.

Rounding:

The value is rounded to 4 digits.

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]SWEEP:STARt<ws><frequency>[units]

[:SOURce:]SWEEP:STARt<ws>MINimum|MAXimum

Examples:

SWEEP:STARt 5KHZ

SWEEP:STARt 5E3

SWEEP:STARt MAXIMUM

SWEEP:STARt MIN

Query
Syntax:

[:SOURce:]SWEEP:STARt?[<ws>MAXimum|MINimum]

Examples:

SWEEP:STARt ?

SWEEP:STARt ? MAX

Response: NR3

4.13.1.10.5 Sweep STOP


This command sets the stop frequency of the sweep:

Arguments
Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Dependent on the frequency range of the current function.

Rounding:

The value is rounded to 4 digits.

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]SWEEP:STOP<ws><frequency>[units]

[:SOURce:]SWEEP:STOP<ws>MINimum|MAXimum

Examples:

SWEEP:STOP 5KHZ

SWEEP:STOP 5E3

SWEEP:STOP MAXIMUM

SWEEP:STOP MIN