Examples:

FSK:RATE 5KHZ

FSK:RATE 5E3

FSK:RATE MAXIMUM

FSK:RATE MIN

Query
Syntax:

[:SOURce]:FSK:RATE ?[<ws>MAXimum|MINimum]

Examples:

FSK:RATE ?

FSK:RATE ? MAX

Response:

NR3

4.13.1.9.5 FSK SOURce


This command selects the FSK modulation source as either internal (then the above settings are effective) or external (and
then the external waveform determines the frequency of modulation).

Arguments
Type:

Character

Options:

INTernal, EXTernal

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:] FSK:SOURce<ws><INT|EXT>

Examples:

FSK:SOUR INT

FSK:SOUR EXT

Query
Syntax:

[:SOURce]:FSK:SOURce?

Response:

INT|EXT

4.13.1.10 Sweep control

The following commands control the sweep functionality:

4.13.1.10.1 Sweep STATe

This command activates or deactivates sweep:

Arguments
Type:

Boolean

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:]SWEEP[:STATe]<ws>ON|1|OFF|0

Examples:

SWEEP:STAT ON

SWEEP OFF

Query
Syntax:

[:SOURce:]SWEEP[:STATe]?

Response:

0|1

4.13.1.10.2 Sweep SPACing

This command sets the sweep spacing as either LINear or LOGarithmic:

Arguments
Type:

Character

Options:

Linear, Logarithmic

Command Type: Setting or Query
Setting
Syntax:

[:SOURce:] SWEEP:SPACing<ws><LIN|LOG>

Examples:

[:SOURce:] SWEEP:SPACing LIN

SWEEP:SPAC LOG

Query
Syntax:

[:SOURce:] SWEEP:SPACing ?

Response:

LIN|LOG