7.13 Connect the OUTPUT connector to the distortion analyzer using a 50 ohm coaxial cable and a

50 ohm feedthrough termination.

7.14

Set the unit to generate a sine waveform with 1KHz frequency and 10Vp-p.

7.15

Set the distortion analyzer to measure distortion in dB and select RMS response.

7.16

CHECK for a reading of more than -65 dB.

5.3.5 Square Transition Times

Specification: < 6 ns (10% to 90%) at 10Vp-p into 50 ohms.

Procedure:

1.

Connect the OUTPUT connector to the oscilloscope input using a 50 ohm coaxial cable and a
50 ohm feedthrough termination.

2.

Select WAVE and SQUARE to generate square waveform with 1MHz frequency and 10Vp-p.

3.

Use the oscilloscope markers to measure the Rise and Fall times of the square, from 10% to

90%

levels.

4.

CHECK for rise and fall times of < 12 ns.

5.3.6 Square Variable Duty Cycle

Specification: variable from 20% to 80% with ± 0.5% accuracy at 1 KHz.

Procedure:

1. Connect the OUTPUT connector to the counter input using a 50 ohm coaxial cable and a 50 ohm

feedthrough termination.

2. Select SQUARE to generate square waveform with 1KHz frequency and 5Vp-p.

3. Set the counter to measure duty cycle and select the automatic trigger mode.

4. On the unit change the symmetry to 20%, 50% and 80%

5. CHECK for duty cycle readings between 19.0% - 21.0%, 49.0% - 51.0% and 79.0% - 81.0%.

5.3.7 Operating Modes

Specification: Continuous, triggered, gated, and burst.

Procedure:

1.

Connect the OUTPUT connector to the oscilloscope vertical input using a 50 ohm coaxial

cable and a 50 ohm feed-through termination.

2.

Push the MODE key on the unit front panel.

3.

Verify that the oscilloscope displays a 1KHz sine wave. Make sure that the Output is ON.