13 marker length, 14 marker state, 15 save – B&K Precision 4078 - Manual User Manual

Page 77

Advertising
background image

Setting
Syntax:

:ARBitrary:MARKer[:ADDRess]<ws><marker address>

Examples:

:ARB:MARK 45

Query
Syntax:

:ARBitrary:MARKer[:ADDRess]?

Example:

:ARB:MARK?

Response:

Marker address in NR1 format.

Considerations: The marker is only output if its address is within the range of addresses currently being run.


4.13.4.13 Marker Length


:ARBitrary:MARKer:LENGth <numeric value>

This command is used to set the marker length. The maximum allowed length of marker is 4000. The marker length
is specified by appending a numeric value, to the MARKer keyword. Maximum MARKER Length 4000 points

Arguments
Type:

Numeric

Command Type: Setting or Query
Setting
Syntax:

:ARBitrary:MARKer:LENGth<ws><length>

Example:

:ARB:MARK:LENG 5

Query
Syntax:

:ARBitrary:MARKer:LENGth?

Response:

NR1

4.13.4.14 Marker State


:ARBitrary:MARKer:STATe <Boolean>


This command is used to enable or disable the marker.

Arguments
Type:

Boolean

Command Type: Setting or Query
Setting
Syntax:

:ARBitrary:MARKer:STATe<ws>ON | OFF (or 1 | 0)0

Query
Syntax:

:ARBitrary:MARKer:STATe?

Response:

0|1


4.13.4.15 Save


:ARBitrary:SAVe


This command is used to save all unsaved arbitrary waveform data into non-volatile memory..

Arguments
Type:

none

Command Type: Setting only
Setting
Syntax:

:ARBitrary: SAVe

Advertising