Fieldbus Communications

mV STANDARD1

40022

Float

2

R

-3000/

3000

Voltage in mV for one standard
sample

mV STANDARD2

40024

Float

2

R

-3000/

3000

Voltage in mV for two standard
samples

mV ZERO

40026

Float

2

R

-3000/

3000

Voltage in mV for the citro sample

mV SAMPLE

40028

Float

2

R

-3000/

3000

Voltage in mV for sample (last
measurement)

mV ACTIVE

40030

Float

2

R

-3000/

3000

Voltage in mV for current sample
(actual mV)

NH4-N VALUE 2

40032

Float

2

R

Measurement value for channel 2
as NH4–N

NH4 VALUE 2

40034

Float

2

R

Measurement value for channel 2
as NH4

NH4-N VALUE 1

40036

Float

2

R

Measurement value for channel 1
as NH4–N

NH4 VALUE 1

40038

Float

2

R

Measurement value for channel 1
as NH4

PROBE P. MIN

40040

Float

2

R

0/2.0

Integrated value of pressure at
filtration probe, if not yet calculated:
nan

PROBE
PRESSURE

40042

Float

2

R

0/2.0

Actual value of pressure at filtration
probe, if not yet calculated: nan

GAIN CORR. 1

40044

Float

2

R/W

0.01/

100.00

Gain correction for channel 1

mV SLOPE

40046

Float

2

R

-3000/

3000

SLOPE of electrode

BUSACTION
ACTIVE

40048

Unsigned

Integer

1

R/W

0/1

Enter 1 to start a BUS action (see
Fieldbus control)

BUS ANALY.START

40049

Unsigned

Integer

1

R/W

0/1

Start the analyzer about the BUS

BUS SERVICE

40050

Unsigned

Integer

1

R/W

0/1

Start the service mode about the
BUS

BUS CLEANING

40051

Unsigned

Integer

1

R/W

0/1

Start the cleaning mode about the
BUS

BUS CALIBRATION

40052

Unsigned

Integer

1

R/W

0/1

Start the calibration mode about the
BUS

BUS CLEAN/CAL.

40053

Unsigned

Integer

1

R/W

0/1

Start the cleaning/calibration mode
about the BUS

BUS PREPUMP
REA.

40054

Unsigned

Integer

1

R/W

0/1

Prepump reagent about the BUS

BUS PREPUMP
CLEA

40055

Unsigned

Integer

1

R/W

0/1

Prepump cleaning solution about
the BUS

BUS PREPUMP
STA.

40056

Unsigned

Integer

1

R/W

0/1

Prepump standards about the BUS

BUS PREPUMP
PRO.

40057

Unsigned

Integer

1

R/W

0/1

Prepump the probe about the BUS

BUS PREPUMP
ALL

40058

Unsigned

Integer

1

R/W

0/1

Prepump all about the BUS

Table 13 Sensor Modbus registers (continued)