A.9 option 5 plumbing and connections, Plumbing and connection options – Hach-Lange AMTAX indoor sc User Manual

Page 88

Advertising
background image

88

Plumbing and Connection Options

A.9 Option 5 plumbing and connections

Option 5 uses an sc analyzer as a 2-channel analyzer with two
FILTRAX (FILTRAX 1 and FILTRAX 2), supplying two continuous
sample streams. The waste from the analyzer and both FILTRAX is
discharged back into the drain through the optional heated drain
hose LZY302 (230V) or LZY303 (115V).

Refer to

Figure 33

and the following instructions for Option 5:

1. Install both FILTRAX into the sample stream. Refer to the

FILTRAX User Manual for more information.

2. Feed the heated hose from FILTRAX 1 through the analyzer

opening (

Figure 33

, item 7). Use Seal Plug #1 to secure.

3. Feed the heated hose from FILTRAX 2 through the analyzer

opening (item 6). Use Seal Plug #1 to secure.

4. Feed the heated drain hose through the analyzer (item 5). Use

Seal Plug #1 to secure.

Note: The two samples lines of the heated drain hose are not used.

5. Connect the heated drain power connections. Refer to

section 3.5.5, page 27

.

6. Connect the heated drain tube to the T-fitting (item 8).

7. Connect the sample line from FILTRAX 1 to the bottom inlet on

Overflow Vessel 1 using the fittings (item 11).

8. Connect the sample line from FILTRAX 2 to the bottom inlet on

Overflow Vessel 2 using the fittings (item 12).

Advertising
This manual is related to the following products: