Plumbing and Connection Options

A.14 Option 9a plumbing and connections

Option 9a uses an sc analyzer as a 2-channel analyzer with two
FILTRAX (FILTRAX 1 and FILTRAX 2). The waste of the analyzer
and both FILTRAX is discharged into an open drain.

Refer to

Figure 38

and the following instructions for Option 9a:

1. Install both FILTRAX into the sample stream. Refer to the

FILTRAX User Manual for more information.

2. Feed the heated hose from FILTRAX 1 through the analyzer

(

Figure 38

, item 6). Use Seal Plug #1 to secure.

3. Feed the heated hose from FILTRAX 2 through the analyzer

(item 5). Use Seal Plug #1 to secure.

4. Feed the drain tube through the analyzer (item 7). Use Seal

Plug #3 to secure.

5. Connect the drain tube to the T-fitting.

6. Connect the sample line from FILTRAX 1 to Overflow Vessel 1

using the fittings. Connect the sample line from FILTRAX 2 to
Overflow Vessel 2 using the fittings.