4 scheduled maintenance, Maintenance – Hach-Lange AMTAX indoor sc User Manual

Page 49

Advertising
background image

49

Maintenance

6.4 Scheduled maintenance

Table 7

lists items, except the electrode, that need to be maintained

by service personnel ONLY. Contact the manufacturer for more
information.

6.5 Replace the membrane cap, electrolyte and electrode

Important Note: Never lubricate the membrane cap or the
electrode with grease, silicone oil or Vaseline. This will cause
damage to the Teflon membrane which will cause poor
performance.

Note: The service life of the Teflon membrane will be reduced by tensides
or organic solvents in the waste water.

For optimal performace, replace the the membrane cap, electrolyte
and electrode regularly (refer to

Table 6, page 48

).

While operating the instrument, the amount of electrolyte should be
between 4 mL and 11 mL. With less amount of electrolyte, the
accuracy will be reduced for low measured values in the related
measuring range. To determine how much electrolyte remains in
the electrode body, refer to

Figure 20

.

Table 7 Repair maintenance items

Description

When to replace

Warranty

Reagent pump for sc analyzer (Valve pumps)

1 year

1 year

Pump head piston pump 10 mL
(Pre-greased cylinder and piston)

1 year

1 year

Switchable Compressor 115/230V

2 years recommended

2 years

Solenoid stirring bar

1 year

1 year

Electrode

Check continuously after 1 year.

Electrode is OK, when slope

with new membrane cap and

electrolyte is in the range of

–57 to –67 mV after 24 h of

membrane cap replacement.

1 year

Figure 20 Determine the electrolyte volume in the electrode body with a measuring tape

Advertising