Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual

Page 118

Advertising
background image

118

Fieldbus Communications

EXHAUST
CONTROL

40272

Unsigned

Integer

1

R/W

0/1

Analyzer checks the exhaust
whether it is blocked; 0=OFF, 1=ON

ELECTROLYTE

40277

Integer

1

R

-32768/3

2767

Left days until next electrolyte
change, negative values shows
overdue

ELECTROLYTE

40278

Unsigned

Integer

1

R/W

0/1

Sets if there is a warning when
electrolyte has to be exchanged;
0=OFF, 1=WARNING

FALSE ELEC DATA

40279

Unsigned

Integer

1

R/W

0/1

Sets if there is an error if the ZERO
value is out of range; 0=OFF, 1=ON

ERROR LIST

40280

Unsigned

Integer

2

R

Errors are coded bit wise. bit
0=TEMP. < 0 °C/ 32°F?, bit
1=ANALYZ. TO COLD, COOLING
FAILED,HUMIDITY ANALY,
HUMIDITY PROBE; PROBE
MISSING, NO HEAT UP,
CUVSENSOR DEFECT,
TEMPSENS DEFECT, CUVHEAT
DEFECT, CUV TOO HOT,
ELECTRODE SLOPE, FALSE
ELEC DATA, MODULES CONTAM.,
DRAIN BLOCKED, SAMPLE1, bit
16= SAMPLE2

WARNING LIST

40282

Unsigned

Integer

2

R

Warnings are coded bit wise, bit 0=
WARMUP PHASE, bit 1=COOLING
DOWN, SERVICE MODE,
REAGENT LEVEL, CLEAN SOLU
LEVEL, ANALYZER TO COLD,
ANALYZER TO WARM, CUV TOO
COOL, MODULES CONTAM.,
STANDARDS LEVEL,
ELECTRODE SLOPE,
reserved,SAMPLE1,
ELEKTROLYTE, bit 14=SAMPLE2

EDIT NAME

40285

String

8

R/W

Name of LOCATION (see menu
system)

ELECTROLYTE

40293

Float

2

R

electrode drift per 24h in mV

Table 13 Sensor Modbus registers (continued)

Tag name

Register #

Data

type

Length

R/W

Discrete

range

Min/Max

range

Description