Troubleshooting, 3 warnings (continued) – Hach-Lange AMTAX indoor sc User Manual

Page 61

Advertising
background image

61

Troubleshooting

REAGENT LEVEL

Continued
measurement

Amount of reagent has
dropped below warning
level set

Check reagent level
and replace if
necessary, then reset
reagent level. The level
is indicated
mathematically and
can only function
reliably if the counter is
only reset when the
solution is changed

On the menu
MAINTENANCE/
TEST/MAINT./
REAGENT

CLEAN SOLU LEVEL

Continued
measurement

Amount of cleaning
solution has dropped
below warning level set

Check cleaning
solution level and
replace if necessary,
then reset cleaning
solution level. The
level is indicated
mathematically and
can only function
reliably if the counter is
only reset when the
solution is changed

On the menu
MAINTENANCE/
MAINT. COUNTER/
CLEANING SOLU.

STANDARDS LEVEL

Continued
measurement

Amount of standard
solution has dropped
below warning level set

Check standard level
and replace if
necessary, then reset
standard level. The
level is indicated
mathematically and
can only function
reliably if the counter is
only reset when the
solution is changed

On the menu
MAINTENANCE/
MAINT. COUNTER/
STANDARD

ELECTROLYTE

Continued
measurement

Elektrolyte level may
be to low or last
CHANGE MEMBRANE
has been more than
90 days ago.

Check amount of
electrolyte, exchange
electrolyte completely if
level is low. Use
process CHANGE
MEMBRANE . DO
NOT add new
electrolyte to old
electrolyte.

Reset warning
manually or automatic
reset after use of
CHANGE
MEMBRANE,
CHANGE
ELECTRODE or
CHANGE
MEASUREMENT
RANGE in menue
TEST/MAINT.

SAMPLE1 / SAMPLE2

Continued
measurement

the amount of sample
is not sufficient
(channel1 / channel2)
This occurs as warning
if SAMPLE
DETECTION is set to
WARNING

Check sample delivery,
make shure that
sample line has no
negative pressure,
check tightness of
piston pump, check
overflow and air valve

Automatic reset when
enough sample is
available or manuell
reset

7.2.3 Warnings (continued)

Warning displayed

Instrument reaction

Cause

Solution

Reset warning