Troubleshooting, 2 error messages (continued) – Hach-Lange AMTAX indoor sc User Manual

Page 59

Advertising
background image

59

Troubleshooting

CUV TOO HOT

Service state, cuvette
heating off!

The cuvette/sample is
over-heated.

Check if incoming
sample is in the
specified range and
increase the cuvette
temperature setpoint
(CONFIGURE>
CUVETTE TEMP.)
Use lowest possible
cuvette temperature.
Call service if error
continue

Reset error manually

ELECTRODE SLOPE

Service state

The error occurs if the
electrode slope is not
in the range of -50mV
to -70mV

Change membrane
and electrolyte, if there
is still a problem, check
standards,
measurement range
and chemistry, check
flow rate. If all OK, but
error still present, fit
new electrode

Reset error manually

FALSE ELEC DATA

Service state

The mV value for Zero
related to Standard 1 is
not within the allowed
range.
Allowed range for
(Uzero-UStandard1)
depending on
measurement range:
Low (0.05 to 20 mg):
5mV to 200mV
Medium (1 to 100 mg):
20mV to 265mV
High (10 to 1000 mg):
50mV to 315mV

Check cleaning
solution (level and
delivery) and sample
delivery, check
Standard 1 (level and if
correct for
measurement range),
change membrane and
elektrolyte. Error may
occure when sample
has very unusual
properties. In this case
the error can be
switched to "OFF" in
the configuration
menue.

Manual or
automatically when the
CHANGE MEMBRANE
or REPLACE
ELCTRODE pocess is
started

MODULES CONTAM.

Continued
measurement

Filter modules heavily
soiled

Clean filter modules
immediately

Reset error manually

DRAIN BLOCKED

Service state

Drain is blocked
(lime?)

Check drain line. Adapt
cleaning interval to
water hardness.

Reset error manually

SAMPLE1 / SAMPLE2

Continued
measurement

the amount of sample
is not sufficient
(channel1 / channel2)
This occurs as error if
SAMPLE DETECTION
is set to ERROR

Check sample delivery,
make sure that sample
line has no negative
pressure, check
tightness of piston
pump, check overflow
and air valve, check
tightness of system.

Automatic reset when
enough sample is
available or manuell
reset

7.2.2 Error messages (continued)

Error displayed

Instrument reaction

Cause

Solution

Reset error

Advertising
This manual is related to the following products: