Troubleshooting, 2 error messages (continued) – Hach-Lange AMTAX indoor sc User Manual

Page 58

Advertising
background image

58

Troubleshooting

ANALYZ. TO COLD

Instrument goes into
the service state

Instrument interior has
been below 4 °C
(39 °F) for more than
5 minutes

Close instrument,
check heating

Reset error manually
TEST/MAINT>RESET
ERROR

NO HEAT UP

Instrument goes into
the service state

Instrument cannot heat
interior adequately.
(internal temperature
< 20 °C (68 °F) for
30 min)

Close instrument,
check heating

Reset error manually

COOLING FAILED

Service state, starts
automatically after
cooling down

Instrument interior is
too warm (> 47 °C
(117 °F) with electrode
temp. 45 °C (113 °F)
> 52 °C (126 °F) with
electrode temp. 50 °C
(122 °F) or >57 °C
(135 °F) with electrode
temp. 55 °C (130 °F) )

Check air filter and
clean/replace,
check fan.

Manual reset or when
temperature falls 2 °C
(3.6 °F) below limit

HUMIDITY ANALY

Service state

There is liquid in the
collecting tray

Identify cause and
rectify

Reset error manually

HUMIDITY PROBE

Service state, the
filtration probe sc is
isolated from the mains

There is liquid in the
filtration probe
enclosure

Immediately take
filtration probe sc out of
operation and contact
service. Take the
filtration probe sc out of
the tank and store the
filter modules so that
they remain moist.
(See operating
instructions for the
filtration probe sc).

Reset error manually

PROBE MISSING

Service state, the
filtration probe sc is
isolated from the mains

The filtration probe sc
is faulty or not
connected

Immediately take
filtration probe sc out of
operation and contact
service. Take the
filtration probe sc out of
the tank and store the
filter modules so that
they remain moist.
(See operating
instructions for the
filtration probe sc).

Reset error manually

TEMPSENS DEFECT

Service state, fan
running, heating off

The temperature
sensor for the internal
instrument temperature
is faulty

Immediately switch off
instrument, contact
service, replace main
circuit board

Reset error manually

CUVSENSOR
DEFECT

Service state, cuvette
heating off

The temperature
sensor for the cuvette
is faulty

Contact service,
replace cuvette/sensor

Reset error manually

CUVHEAT DEFECT

Continued
measurement

The cuvette is not
being heated
adequately

Close instrument door
and wait 10 minutes if
error accurs again,
contact service

Reset error manually

7.2.2 Error messages (continued)

Error displayed

Instrument reaction

Cause

Solution

Reset error