2 sensor setup menu (continued), Operation – Hach-Lange AMTAX indoor sc User Manual

Page 40

Advertising
background image

40

Operation

MAINTENANCE (continued)

PROCESS

Information what instrument is currently doing (measurement, calibration etc.)

REMAINING TIME

Remaining time for current process, counting down to zero

LIST OF VALUES

List of the last 10 measured values

MAINT. COUNTER

Counter for reagent and consumables

OPERATING HOURS

Indicates the instrument's operating hours.

REAGENT

Displays the current level of the reagent.

CLEANING SOL.

Displays the current level of the cleaning solution.

STANDARD SOL.

Displays the current level of the standard.

REPLACE ELECTRO.

Date of the last electrode change

CHANGE MEMBRAN

Last electrode membrane and electrolyte change.

AIR FILTER PADS

Days left until the next air filter change/clean.

PISTON PUMP

Days left until the next pump piston and cylinder replacement (AMTAX piston pump)

REAGENT PUMP

Number of pump strokes performed by the reagent metering pump

CLEANING PUMP

Number of pump strokes performed by the cleaning metering pump

STATUS MODULES

Only if filtration probe is registered: indicates the state of the modules.

CLEANING MODULS

Only if filtration probe is registered: last filter module cleaning.

NEW MODULES

Only if filtration probe is registered: last filter module replacement.

PUMP MEMBRANE

Only if filtration probe is registered: date of the last pump membrane replacement (filtration
probe sample pump).

COMPRESSOR

Only if filtration probe is registered: days left until the replacement of the air compressor.

ELECTROLYTE

counts down from 90 days. Negative values indicate that electrolyte change is overdue. Reset
by "CHANGE MEMBRANE" or "CHANGE ELECTRODE" process

TEST/MAINT

Maintenance processes

SIGNALS

PROCESS

Indication of what the instrument is doing.

REMAINING TIME

Indication of the time left for the currently ongoing process

mV ACTIVE

Current electrode potential (For electrode data ranges refer to

Table 10, page 62

).

CUVETTE TEMP.

Current measuring cell temperature

ENCLOSURE TEMP

Current temperature in the instrument

COOLING

Current speed of the enclosure fan in %

HEATING

Current enclosure heating power

PRESSURE ANALY

Current pressure in the metering system of the valve block in mbar

HUMIDITY ANALY

Indication of whether there is liquid in the collecting tray

STATUS MODULES

Only if filtration probe is registered: shows status of filtration modules (0–100 %)

PRESSURE P. MIN

Only if filtration probe is registered :shows the averaged minimum pressure at the filter
modules

PRESSURE PROBE

Only if filtration probe is registered: shows the actual minimum pressure at the filter modules

PROBE HEATING

Only if filtration probe is registered: indication of switching state of sample tubing heating

HUMIDITY PROBE

Only if filtration probe is registered: indication of whether there is moisture in the probe
enclosure

DRAIN HEATING

only in 1 or 2 channel mode: shows status of drain heating

PROCESS

Indication of what the instrument is doing.

REMAINING TIME

Indication of the time left for the currently ongoing process

SERVICE MODE

Instrument can be adjusted to service mode (system free of liquids, thermal management and
compressor for filter probe (if installed) active

5.2 Sensor setup menu (continued)