Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual

Page 115

Advertising
background image

115

Fieldbus Communications

SET OUTMODE
CAL.

40086

Unsigned

Integer

1

R/W

0/1

Set output mode for calibration;
0=HOLD, 1= TRANSFER VALUE

DISCHARGE
CLEAN.

40087

Unsigned

Integer

1

R/W

0/10

Discharged values after a cleaning

SET OUTMODE
CLE.

40088

Unsigned

Integer

1

R/W

0/1

Set output mode for cleaning;
0=HOLD, 1= TRANSFER VALUE

SET OUTMODE
SER.

40089

Unsigned

Integer

1

R/W

0/1

Set output mode for service mode;
0=HOLD, 1= TRANSFER VALUE

LOCATION2

40090

String

8

R/W

Location for the measuring channel
two for the sample

SET PARAMETER 2

40098

Unsigned

Integer

1

R/W

19/42

Select the parameter for channel 2;
19=NH4-N, 42=NH4

GAIN CORR. 2

40099

Float

2

R/W

0.01/100

.00

Gain correction for channel two

MEAS.UNITS 2

40101

Unsigned

Integer

1

R/W

0/2

Measurement units for channel 2;
0=mg/L, 2=ppm

HUMIDITY ANALY

40102

Unsigned

Integer

1

R

0/100

Humidity analyzer in percent

SOFTWARE
PROBE

40103

Float

2

R

0/3.4028

2347E+

38

Software version of the filtration
probe

HUMIDITY PROBE

40105

Unsigned

Integer

1

R

0/100

Humidity filtration probe in percent

PROCESS STATE

40107

Unsigned

Integer

1

R

0/1/2/3/4/
5/6/7/8/9/

10/11/12/

13/14/15/
16/17/18/

19/20

Process status is coded as enum

list; enum value 0=service mode,
enum value 1=citrocal in proc, cal1
in proc., cal2 in proc., measuring
1…, interval, initialisation, serv.in
proc., cleaning, warm up phase,
measuring 2…, prepump reag.,
prepump clean.,prepump standard,
prepumping probe, flushing, start by
BUS, warming up, reserved,
prepump.sample; enum value
20=validation

LAST CALIBRAT.

40108

Time2

2

R

Date of the last calibration

START BY BUS

40110

Unsigned

Integer

1

R/W

0/1

To set the analyzer to START BY
BUS mode (see Start by BUS)

FIELDBUS

40111

Unsigned

Integer

1

R/W

0/2

To trigger a measurement serie in
START BY BUS mode (see Start by
BUS)

NUMBER OF
MEAS.

40112

Unsigned

Integer

1

R/W

1/100

Number of measurements in a
START BY BUS measurement
series (see Start by BUS)

DISCHARGE BUS

40113

Unsigned

Integer

1

R/W

3/10

Discharge values at the beginning
of a START BY BUS series

AVERAGE

40114

Unsigned

Integer

1

R/W

Number of measurement values
that result in an average value in a
START BY BUS measurement
series.

Table 13 Sensor Modbus registers (continued)

Tag name

Register #

Data

type

Length

R/W

Discrete

range

Min/Max

range

Description