2 sensor setup menu (continued), Operation – Hach-Lange AMTAX indoor sc User Manual

Page 39

Advertising
background image

39

Operation

CONFIGURE (continued)

SAMPLE DETECTION

OFF/WARNING/ERROR

Determines instruments reaction when amount of available sample is too low. When
instrument is in filter probe mode, deactivating the sample detection will switch a deactivated
"STATUS MODUL.ERR" to 14 %

EXHAUST CONTROL

ON/OFF

Determines instrument reaction when drain is blocked

ELECTROLYTE

WARNING/OFF

Determines whether a warning is initiated when electrolyte level is low/ last change of
membrane is more than 90 days ago.

FALSE ELEC DATA

ERROR/OFF

Determines whether an error is initiated when the mV-Zero value of the electrode is not within
a certain range (see list of errors)

DEFAULT SETUP

Resets factor and settings to factory settings.

LAST CHANGE

Indication of the last change of a setting in the configuration menu.

MAINTENANCE

INFORMATION

LOCATION 1

Indication of measuring location 1

LOCATION 2

Indication of measuring location 2 on two-channel version

TYPE

Indication of instrument type

SENSOR NAME

Indication of instrument name

SERIAL NUMBER

Indication of serial number

RANGE

Indication of measuring range

OPTION

Indication of instrument option (filtration probe/1-channel/2-channel)

SOFTWARE PROBE

Filtration probe software

SOFTW. AMTAX

Instrument software

LOADER

Detailed information on the software in the instrument

APPL

Detailed information on the software in the instrument

STRUCTURE

Detailed information on the software in the instrument

FIRMWARE

Detailed information on the software in the instrument

CONTENT

Detailed information on the software in the instrument

LANGUAGE

List of languages that are supported by the installed language package.

CALIB. DATA

LOCATION1

Indication of measuring location 1

GAIN CORR

Indicates the correction factor set for correcting the measured values at measuring location 1.

DATE

Indicates the date of the last change to the correction factor.

LOCATION2

Available with the 2-channel version.

GAIN CORR

Indicates the correction factor set for correcting the measured values at measuring location 2.

DATE

Indicates the date of the last change to the correction factor.

mV ZERO

Electrode signal at the zero point

mV STANDARD 1

Electrode signal with standard 1

mV STANDARD 2

Electrode signal with standard 2

mV SLOPE

Electrode signal change per decade

LAST CALIBRAT.

Time of the last calibration

mV ACTIVE

Current electrode potential

5.2 Sensor setup menu (continued)