Operation

MAINTENANCE (continued)

SET OUTMODE

Value that is output in the service state. HOLD = last measured value,
SET TRANSFER = Transfer value that is programmed on SC controller

START

Leave service mode, start measurement

REAGENT

Resets the maintenance counter after reagent change

CLEANING SOLU.

Resets the maintenance counter after changing the cleaning solution.

STANDARD SOL.

Resets the maintenance counter after changing the standards.

AIR FILTER PADS

Menu-based process for changing the air filter pads, resetting the maintenance counter

CHANGE MEMBRAN

Menu-based process for membrane-cap change of the electrode, sets new date in the
maintenance counter and resets electrolyte counter

REPLACE ELECTRODE Menu-based process for electrode change, sets new date in the maintenance counter

PISTON PUMP

Days left until the next pump piston and cylinder replacement (AMTAX piston pump), reset
after replacement of the pump

REAGENT PUMP

Number of pump strokes performed by the reagent metering pump, reset after replacement of
the pump

CLEANING PUMP

Number of pump strokes performed by the cleaning metering pump, reset after replacement
of the pump

PREPUMPING

PREPUMP ALL

All liquids are pre-pumped in succession.

PREPUMP REAG.

The reagent is pre-pumped.

PREPUMP CLEAN.

The cleaning solution is pre-pumped.

PREPUMP
STANDARD

The standards are pre-pumped.

PREPUMPING
PROBE

Only if filtration probe is registered: The filtration probe and modules are bled and
pre-pumped.

PREPUMP SAMPLE

Only if filtration probe is registered: sample is pumped from the filtration probe for 1 minute

MODULE CLEAN.

Menu-based process for cleaning the filter modules, automatically resetting the maintenance
counter. With filtration probe.

NEW MODULES

Only if filtration probe is registered: last filter module replacement.

PUMP MEMBRANE

Only if filtration probe is registered: shows the days left until the next change of pump
membrane (Filter Probe sc sample pump), Rest of the counter.

COMPRESSOR

Only if filtration probe is registered: days left until the replacement of the air compressor.
Reset of the counter.

CLEANING

Trigger an automatic cleaning, then start measurement

FLUSHING

Pumps all liquids in succession. Put all tubings that go to reagents, standards and cleaning
solutions into deionised water and start FLUSHING prior to taking instrument out of operation

RESET ERROR

Reset all error messages

CHANGE RANGE

Software changes to another measuring range: ATTENTION, the appropriate standards must
be used!

UPDATE PROBE

Enables the filtration probe software to be updated.

Elec. Type Change

Not used at the moment

FIELDBUS

ENABLED/DISABLED: Enable external control of instrument by Fieldbus. NOTE: When
instrument is put into SERVICE MODE by menu, the Fieldbus control is temporarily disabled.

OPTION

Sets the instrument to filter-probe/ 1channel/ 2channel mode. Switching the options needs
modification of hardware!

VALIDATION

Menu based process to measure external samples. When "Modification required" is
displayed: Disconnect sample tubing from overflow vessel, plug overflow vessel and put
sample tubing into external sample. After process: Unplug overflow vessel and reconnect
sample tube.