3 warnings, Troubleshooting 7.2.3 warnings – Hach-Lange AMTAX indoor sc User Manual

Page 60

Advertising
background image

60

Troubleshooting

7.2.3 Warnings

Warning displayed

Instrument reaction

Cause

Solution

Reset warning

WARMUP PHASE

Instrument is warming
up sample tubing after
start (de-frosting)

If there is a risk that the
sample tubes are
frozen, a warning is
displayed

As far as possible, wait
until end of the
warm-up phase
(except if certain there
is no frost), to cancel
place instrument in
service state and start
the measurement
again

Automatic

COOLING DOWN

Fan 100 % , standstill
until cool enough

Instrument is cooling
down after start using
ventilation, if it was
heated excessively

Wait until instrument
has cooled down
enough

Reset, automatically as
soon as cool

ANALYZER TO COLD

Measurement

Interior of instrument is
cooler than 15 °C
(59 °F)

Close instrument door,
if necessary check
heating

Reset, automatically as
soon as warmer

ANALYZER TO WARM

Measurement, but no
more air cleaning

At very high internal
temperature, the air
cleaning of the filter
modules is disabled so
as to generate less
heat. internal temp =
target temp of cuvette

Change/clean air filter,
check air ducts for
blockage, check
enclosure fan, is
ambient temperature
allowed? If necessary
increase electrode
temperature

Reset, automatically as
soon as cool

CUV TOO COOL

Continued
measurement

The cuvette is not
being heated
adequately 2 min after
sample change:
temp. = ((target temp.
of cuvette) –1 °C
(34 °F))

Close instrument door,
check/fit cuvette
insulation. Attention: If
this warning occurs
during the calibration,
there is a risk of
incorrect
measurements, for this
reason close door
during calibration!

Automatic

ELECTRODE SLOPE

Continued
measurement

The warning occurs if
the electrode slope is
not in the range of
–55mV to –67mV

Change membrane
and electrolyte, if there
is still a problem, check
standards and
chemistry, check flow
rate. If all OK, but
warning still present, fit
new electrode

Automatic

MODULES CONTAM.

Continued
measurement

Filter modules soiled

Clean filter modules
soon

Automatic

SERVICE MODE

Service state

The instrument is in the
service state or is
switching to this state

Automatic when the
service state is left