Troubleshooting – Hach-Lange AMTAX indoor sc User Manual

Page 63

Advertising
background image

63

Troubleshooting

Refer to

Table 12

for additional errors messages.

The electrode slope is between –60 and
–65 mV and the 3 calibration values are
all significantly offset to the positive
range.

- The electrolyte is almost empty
- electrode body leaking
- Membrane cap leaking

- Electrolyte almost completely
evaporated after extended operation
- Tighten membrane cap
- Leaking electrode body: order new
electrode

All mV values for the calibration data
only differ in the figures after decimal
separator (almost constant).

- Amplifier board faulty

- Fit new amplifier board

Table 11 Error messages (continued)

Error description

Diagnostics

Solution

Table 12 Additional error messages

Error description

Diagnostics

Solution

The calibration values

fluctuate

excessively

- Electrode faulty

- Replace electrolyte and membrane
cap

The slope is between –40 and –45 mV
after the calibration

- The measuring range 0.05–20 mg/L
NH

4

–N is adjusted and the wrong

standard solutions 10 and 50 mg/L
NH

4

–N are used.

- Insert the right standard solutions and
complete the procedure for changing the
measuring range in the service menu.

The slope decreases, the membrane
can be damaged. The values of the
electrode slope can decrease after
1–2 weeks to –40 mV and –50 mV.

- Membrane is damaged (for example
with silicone oil).

- Remove the measurement chamber
completley
- Clean it thoroughly

A large continuous drift is on the
measured values (up to 2 mg in 24 h).

- Electrode cap is damaged. The
electrolyte crystallize on the cover/cap
and evaporate very quickly.

- Re-new electrolyte
- Insert new electrode

Fluctuations of measured values and
interference of the calibration in the
lower measuring range.

-

Insufficient sample when sample is

required for the determination of the

zero value

- Make sure that sufficient sample is
available

All three electrode values increase
within 24 h over 8 mV to positive values.

- The ectrolyte level dropped out below
4 mL.

- Empty the electrode body, rinse it with
distilled water and fill it with new
electrolyte.

Advertising
This manual is related to the following products: