A.16 option 10a plumbing and connections, Plumbing and connection options – Hach-Lange AMTAX indoor sc User Manual

Page 102

Advertising
background image

102

Plumbing and Connection Options

A.16 Option 10a plumbing and connections

Option 10a uses an sc analyzer with any type of sample
preparation that delivers a continuous sample stream that cannot
be pressurized. The waste of the analyzer is discharged into an
open drain.

Refer to

Figure 40

and the following instructions for Option 10a:

1. Install the sample preparation unit.

2. Feed the sample line from the sample preparation unit through

the analyzer (

Figure 40

, item 2). Use Seal Plug #3 to secure.

3. Feed the drain tube through the analyzer (item 3). Use Seal

Plug #3 to secure.

Note: Tubes can be pushed through prepared holes on Seal Plug #3.

4. Seal any unused holes with Seal Plug #3.

5. Connect the drain tube to the T-fitting.

6. Connect the sample line from the sample preparation to the

bottom inlet on the overflow vessel using the fittings.

Advertising
This manual is related to the following products: