B.4 modbus register information, Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual

Page 112

Advertising
background image

112

Fieldbus Communications

Important Note: Do not try to change the listed register addresses, otherwise the
instrument may malfunction or become inoperable.

The Fieldbus register contains FFFFh (65536dec) when the feature is disabled.

A measurement series is initiated with entering "1" to register 40111 (Enter "2" for
2-channel instruments to start measurements on channel 2). The register will return to "0"
after the measurement series is done. The measurement results can be found at 40001
(channel 1) and 40165 (channel 2).

A value will appear every AVERAGE and at the end of a series if remind measurement(s)
exist. Example: NUMBER OF MEAS is set to 5 and AVERAGE to 2. The result are 3
values, the first is the average of measurement 1 and 2, the second is the average of
value 3 and 4, the last value is the reminding value of the 5th measurement

Note: Internal processes like calibration and cleaning will be interrupted from a measurement
series.The interrupted process will start after the end of the measurement series. To use the START
BY BUS feature, the sample has to be available for calibration, cleaning and rinsing purposes. An
ongoing measurement series will not be interrupted by internal processes.

B.3 External trigger contact, Control by external signal

If the controller board is equipped with an external input terminal (optional board version),
measurements can be issued by applying an external DC voltage of 15V to 30V to the
terminal for longer than 3 seconds. When the field bus control is activated, the input will
issue a forced measurement as described within Fieldbus control.

When the START BY BUS feature is activated, the external input will issue a
measurement series as described within the START BY BUS section.

Note: Only measurements on channel 1 can be started with the external contact on 2 channel
instruments.

B.4 Modbus register information

Table 13 Sensor Modbus registers

Tag name

Register #

Data

type

Length

R/W

Discrete

range

Min/Max

range

Description

MEASURE VALUE 1

40001

Float

2

R

Actual measurement value from
channel one

LOCATION1

40005

String

8

R/W

Name of LOCATION 1 (see menu
system)

MEAS.UNITS 1

40013

Unsigned

Integer

1

R/W

0/2

Measurement units for channel 1;
0=mg/L, 2=ppm

CUVETTE TEMP.

40014

Float

2

R

-50/

99.99

Actual cuvette temperature in °C

CALIB.INTERVAL

40016

Unsigned

Integer

1

R/W

0/1/2/3/4

Calibration interval; 0=OFF,
1=12h,2=24h, 3=36h, 4=48h

CALIB.START

40017

Unsigned

Integer

1

R/W

0/1/2/3/4/
5/6/7/8/9/

10/11/12/

13/14/15/
16/17/18/
19/20/21/

22/23

Acceleration time for calibration
(24 hour format) 0=0 o’clock to
23=23 o’clock

ENCLOSURE
TEMP

40020

Float

2

R

-50/

99.9

Temperature inside the analyzer