Installation

3.5.3 Insert tubing and/or cables

3.5.4 Connect the Filter Probe to the analyzer

CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.

DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the analyzer.

Important Note: Make sure that for the use of a filter probe, the
level of the water in which the filtration probe is submersed, is
below the level of the analyzer.

1. Open the enclosure door and secure with the door hook.

2. Open the analyzer panel.

3. Remove the two screws from the protective cover and remove

the cover (item 1,

Figure 14, page 24

).

4. Connect the earth ground (green/yellow) wire (item 9,

Figure 16, page 28

) from the Filter Probe sc to the ground

terminal (item 5,

Figure 16, page 28

).

5. Connect the power connector to the appropriate terminal

connection (items 4 and 11,

Figure 16, page 28

).

6. Remove the three screws securing the bottom panel cover.

(item 2,

Figure 14, page 24

). Remove the panel.

7. Connect the data connector (item 10,

Figure 16, page 28

) to

the main board (item 12,

Figure 16, page 28

).

8. Install all covers and panels.

9. Connect the white air tubing (item 8,

Figure 16, page 28

) from

the Filter Probe sc to the air tubing connection on the analyzer
(

Figure 16, page 28

).

10. For sample and drain line connections refer to

A.4 , page 79

.

1

Guide the tubing or cables through
the enclosure openings
(

Figure 15

).

2

Push the plug from the top onto the
tubing or onto the cable.

3

Pull down the plug with the tubing
or the cables. Seal any unused
entry with Plug #3.