A.12 option 8a plumbing and connections, Plumbing and connection options – Hach-Lange AMTAX indoor sc User Manual

Page 94

Advertising
background image

94

Plumbing and Connection Options

A.12 Option 8a plumbing and connections

Option 8a uses an sc analyzer with the FILTRAX. The waste of the
analyzer is discharged back into an open drain.

Refer to

Figure 36

and the following instructions for Option 8a:

1. Install the FILTRAX into the sample stream. Refer to the

FILTRAX User Manual for more information.

2. Feed the heated hose from the FILTRAX through the analyzer

opening (

Figure 36

, item 3). Use Seal Plug #1 to secure.

3. Feed the drain tube through the analyzer opening (item 5).

Use Seal Plug #3 to secure.

Note: Tubes can be pushed through prepared holes on Seal Plug #3.

4. Connect the drain tube to the T-fitting.

5. Connect the FILTRAX sample line to the bottom inlet on the

overflow vessel using the fittings (item 4).

6. Feed the drain tube to a lower drain (maximum 2 m/6.5 ft).

Advertising
This manual is related to the following products: