Installation

Locally supplied
The following must be provided locally by the user:

3.3 Positioning

Hach Lange recommends installing the instrument on a sturdy laboratory table that can more
than support the 55kg weight. Place it on a clean flat surface, and convenient to the power
source. The display screen should be at head-height for easy viewing and operation.

3.4 Installation

3.4.1 Pre-installation

You will need the antifoam cartridge filled with the antifoam liquid as part of the installation
process. Prepare this before starting the installation using the instructions in

Antifoam cartridge

preparation on page 23

.

In addition make sure you have suitable standards available before performing any calibration,
see

Standards requirements on page 45

for calibrating the instrument sensors.

Your A1100 electrochemical sensor has been thoroughly cleaned and tested at the factory
before shipment and installed in the instrument. It has been shipped with a cartridge containing
a membrane and electrolyte pre-installed to protect the sensor head. The sensor must be
removed from the instrument and the cartridge removed and replaced with a new one prior to
first use to make it fully operational. These steps are described in the A1100 cleaning
procedure for 6110 TPA

document supplied with the instrument. The new cartridge is included

in the sensor recharge kit, also supplied with the instrument.

Solenoid valve 2/2 NC with O-ring

33122

Filter with Goretex membrane (52 items)

33124

Piercing tip x1 (for PET)

DG33308

Piercing tip x2 (for metal)

33127

Piercing tip mounting tool

33159

Pneumatic tubing kit

33154

Antifoam recharge bottle 1L

33156

Antifoam cartridge (empty)

33157

Antifoam priming beaker

33161

Syringe for antifoam cartridge

33162

Sensor recharge kit

2956A-A

Outlet sample tube assembly

DG33267

USB key

60930

USB client cable including connectors, length = 3m

32533.03

Sensor support (part of A1100-S00T)

32205

PPS sensor storage and calibration cap (part of A1100-S00T)

28129

Purge gas for O

2

and CO

2

sensor (CO

2

)

6 to 7 bar (87 to 102 psia), purity > 99.9%

Forcing gas for piercing device (Air or N

2

)

5.5 to 6 bar (80 to 87 psia)

Reference gas for CO

2

sensor (optional)

5.5 to 6 bar (80 to 87 psia)

Loctite

®

222 sealant (or similar)

For locking and sealing piercing tips

Precision certified barometer

For calibrating the barometric pressure sensor

Precision certified pressure gauge (0-7 bar)

For calibrating the pressure sensor

CAUTION

The instrument is heavy (55 kg) so extreme care must be taken when handling, to avoid
damaging the instrument or inflicting personal injury. It is highly recommended to use a
pulley system (or similar) attached to the lifting ring located on the top of the instrument when
moving it.