2 warning messages, Troubleshooting 10.1.2 warning messages – Hach-Lange ORBISPHERE 6110 User Manual User Manual

Page 66

Advertising
background image

64

Troubleshooting

10.1.2 Warning messages

Important Note:

The value in the right hand column of the above table is a direct correlation to

the value in the warning flag column in the measurement file Result.txt. If a value in the warning
flag column does not exist in the above table it is because it is a composite code meaning more
than one warning is relevant (e.g. a value of 6 in the warning flag column equates to warning
messages 4 and 2 from the table above). Be aware that this is a hexadecimal code meaning that
if warning codes 8 and 2 from the above table are relevant, it will be displayed as an A in the
warning flag column of the measurement file. If you have any difficulties relating the
measurement file warning flags to the codes in the above table, please contact your Hach Lange
representative for assistance.

Warning

code

Displayed message

Probable cause

Solution

Code in

results

file

HEX001

W1: Unstable reference
pressure.

The variation of reference
pressure is too high.

Check the pressure stability of
the input reference gas.

1

HEX002

W2: Unmatched package size.

The measured package height is
different from the value defined in
the package definition.

Check the package size
definition.

2

HEX004

W4: Measurement timeout.
Accuracy of dissolved O2 not
guaranteed.

The measurement pressure is
lower than 3.5 bar.

Check the black seal on the
piercing unit.

4

Check the purge gas pressure.
Check the forcing gas pressure.
Check the Goretex filter.

Gas path not airtight.

Check the airtightness of the gas
path between the piercing unit
and the measurement chamber.

HEX010

N/A

A spontaneous and/or too fast
degassing has occurred.

Restart the measurement with
another package.

10

The package (especially in the
case of cans) is losing pressure
due to a leak.

Verify the package is leak free i.e.
watertight and airtight

The sensor has become
unstable.

Perform an oxygen sensor
maintenance procedure

HEX020

W20: Measurement process
incomplete. Insufficient initial
package pressure.

The pressure in the package is
too low to continue with a
measurement.

Restart the measurement with
another package.

20

HEX040

W40: Spontaneous degassing...

A spontaneous degassing is
detected at the end of the first
expansion.

Restart the measurement with
another package.

40

HEX080

W80: Measurement process
incomplete. Insufficient package
pressure during ultrasound
stage.

The measured pressure is too
low to compute the second phase
of the measurement.

Check foam intrusion in the gas
path.

80

Check Goretex filter.

HEX100

N/A

Possible problem with instrument
configuration

Contact your Hach Lange
representative to analyze the
problem.

100

HEX200

N/A

Possible problem with instrument
configuration

Contact your Hach Lange
representative to analyze the
problem.

200

HEX400

N/A

True CO2 may not be reliable
due to CO2 purge gas purity out
of specification.

Check gas cylinder purity, tubing
and connections to 6110 and
recalibrate CO2 sensor

400