10.1.3 Red LED’s on startup screen

Item

Possible cause

Corrective action

Boards

Bad communication between boards. This
may happen when switching the instrument
on.

Switch instrument off, wait 10 seconds and switch back
on. If the LED is still red call Hach Lange service support.

O

2

sensor

No communication between sensor and
board.

Check connection of the oxygen sensor. If problem
persists, call Hach Lange service support.

O

2

sensor residual

too high

O

2

concentration is not below residual

threshold (0.1 mbar)

Wait until a residual < 0.1 mbar is reached.
Check gas purity and gas connections.
Perform an O

2

sensor verification.

Check sensor protection cap positioning.
If required, perform an O

2

sensor maintenance (necessary

for the initial start up).
Note:

A high residual value will also affect the TC sensor

readings and may cause them to go out of bounds. The
problem with the TC sensor readings should self-correct as
the residual value returns to normal.

TC sensor

No communication between sensor and
board. Sensor damaged.

Switch instrument off, wait 10 seconds and switch back
on. If the problem persists call Hach Lange service
support.

Temperature
sensor

No communication between sensor and
board. Sensor damaged.

Switch instrument off, wait 10 seconds and switch back
on. If the problem persists call Hach Lange service
support.

Flow sensor

No communication between sensor and
board. Sensor damaged.

Switch instrument off, wait 10 seconds and switch back
on. If the problem persists call Hach Lange service
support.

Pressure sensor

No communication between sensor and
board. Sensor damaged.

Switch instrument off, wait 10 seconds and switch back
on. If the problem persists call Hach Lange service
support.

Foam detector

Ingress of foam or liquid in the sampling
tube.

Activate the needle and the purge valves in the Global
System View screen (see

Global system view on

page 62

).

Electrovalves

Switch instrument off, wait 10 seconds and switch back
on. If the problem persists call Hach Lange service
support.

Motor

Bad initialization or communication.

Initialize motor position in the Actuator Verification screen
(see

Actuator verification on page 60

).

Ultrasound

Sonotrode failure

Activate the emergency STOP button and call Hach Lange
service support.

Protection cover

Cover is not on the instrument or is badly
positioned.
Cover detection sensor failure.

Check that the cover is correctly in place and fixed with
screws.
Check that the position detector is in place (in the back of
the instrument).

Emergency STOP

The emergency STOP button is in the
activated position (pushed in).

This can happen during transportation. Release the button
by turning it clockwise.