Installation

7.

Attach the end of the second piece of tubing to the antifoam cartridge by pressing until the
connector clicks into place on the inlet valve. Then, push on the end of the syringe to empty
the contents into the cartridge. The syringe has a volume of 100mL and the cartridge a
volume of 280mL, so repeat the process until the antifoam cartridge is full.

8.

On completion, rinse out the syringe and all tubing with water. Screw the cap back onto the
antifoam container and store the tools and container in a cool dry place.

Important Note:

After installation of the antifoam cartridge in the analyzer, it is critical that any

air bubbles be cleared from the antifoam circuit before taking any measurements. To do this,
select the Antifoam system option from the Actuator verification option in the Maintenance
menu and follow the instructions as described in

Antifoam system on page 61

. Select 50

injections and stop the process when the system constantly spits a small stream of antifoam at
each stroke denoting the antifoam circuit is free of air bubbles.