Configuration

5.3 Package parameters

5.3.1 Package management

Package management
Add new package definitions (up to a maximum of 100), and edit or delete existing definitions. To edit or
delete a package, first select the package from the list displayed by touching the package name on the
screen, and then press either Edit or Delete as appropriate.

New

Create a new package. The new package will be created using the default package
parameters. You will then need to select the new package and edit these default parameters
to define the new package parameters.

Delete

Confirm the deletion to delete the package from the list

Edit

A new screen is displayed to allow you to change the definition of the package. Select each
tab in turn (detailed below) to define the package definitions.

Product tab

Solubility

Enter the type of beverage in the package.

Headspace

Check this box for a fast measurement sequence that provides data from the
headspace only.

Equilibrated

This is only available if Headspace (previous box) is checked. Check this box if
you require the additional total O

2

and dissolved O

2

measurements taken from

the headspace.

Slow
Decompress

If checked, then after the measurement process has completed, a slow release of
any remaining pressure will take place so the package can be handled safely.

Formula

Define up to five different formulae for computing a new parameter. The formula
definitions must already have been entered into the system (see

Formula

management on page 36

for details).

Package tab

Comment

Enter a free-format alphanumeric text. This text will appear against the package
description in the main package management screen.

Brimful volume

Enter the volume of the package when full to overflowing.

Height

Enter the overall height of the package. This information is used if you activate
the Package size check option in

Instrument parameters on page 37

.

Tilt position

Enter the tilt position of the package holder (optional). This should be 1, 2, or 3
and corresponds to the number on the tilt knob on the front of the instrument (

No.

4

in

Figure 2 on page 15

).

Pck backstop

Position of the package backstop (optional). The backstop is calibrated on a scale
of zero to seven.

Package type

Select the correct image for the type of package being measured (bottle, can,
etc.). The setups for each icon are different, so it is important to select the correct
one.

Note:

The above reference values are also displayed at the start of the measurement process for that

package, so the operator can adjust the instrument accordingly (tilt position, etc.).

Alarms tab

Alarm
parameters

Select the parameter to define the alarm settings, and enter the minimum and
maximum values allowable for each parameter listed. When the measurement
value falls below the minimum level or exceeds the maximum level for any
parameter listed, an alarm will be triggered.

Factors tab

Factors

Small adjustments can be made to the measurements listed by entering a factor.
This defaults to 1.000 for all measurements to indicate no adjustment. If this is
changed then the measurement calculated by the instrument will be multiplied by
this factor to give an adjusted measurement value.