Installation

3.4.2 Physical installation

An installation wizard will start automatically once power is connected to the instrument for the
first time.

1.

Connect the instrument to a power source - see

Power supply on page 23

.

2.

Turn the instrument ON. An instrument auto check routine will start automatically. As this is
the first time the instrument will have been powered up, the O2 sensor residual and the
Antifoam level

LEDs will always be red. This is normal as the system has not yet been fully

commissioned and these can be ignored at this stage.

3.

The wizard will start automatically. Press the Installation button to proceed.

4.

At the end of the process select the auto-check option. See

Instrument startup on page 30

for a more detailed explanation of the auto-check process and results.

3.4.3 Post-installation procedures

1.

The instrument has been delivered with English as the default language. However, a

USB

memory stick

is supplied with other available languages (German, Spanish, Chinese, and

Japanese). If you wish to keep English as the preferred language please proceed to the
next step (Step 2 below), otherwise follow these instructions for installing a new language:

a.

Make sure that ActiveSync® is installed on your PC. If you do not have it installed on
your PC, please refer to

Microsoft ActiveSync® configuration on page 22

for installation

details.

b.

Connect your PC to the instrument via the USB client connector (No. 5 in

Figure 3 on

page 16

). ActiveSync® will start automatically. By default ActiveSync® proposes to set

up a partnership with the instrument. This is not required, so select NO before
continuing.

c.

Insert the supplied

USB memory stick

into your PC.

d.

Inside the “Setup” folder, double click on the “setup.exe” executable file. Follow the
on-screen instructions on your PC and then on the instrument to install the preferred
language software.

2.

Set the security levels, user ID’s and passwords as described in

Security and user

management on page 38

.

3.

Configure the instrument parameters as described in

Configuration on page 33

.

4.

Perform a barometric sensor calibration as described in

Barometric pressure sensor on

page 46

.

5.

Perform an O

2

sensor calibration as described in

Oxygen sensor on page 48

. Wait until the

residual value of the sensor is low enough (controlled automatically by the instrument). This
operation may take around 1-2 hours depending on storage conditions.

Note:

The pressure, flow, temperature and CO

2

sensors are more stable and will not require calibration

prior to use.

The instrument should now be ready for use.

If a problem should arise with the instrument, please refer initially to

Troubleshooting on

page 63

. If the problem cannot be overcome, please contact your Hach Lange representative

who will be happy to assist you.