2 power supply, 3 emergency stop button, 4 antifoam cartridge preparation – Hach-Lange ORBISPHERE 6110 User Manual User Manual

Page 25: Antifoam cartridge, Installation 3.5.2 power supply

Advertising
background image

23

Installation

3.5.2 Power supply

Important Note:

In order to get optimum performance from the instrument always keep the gas

and power supplies switched on.

3.5.3 Emergency STOP button

This is located on the right side of the instrument (No. 5 in

Figure 4 on page 17

). Should the

need arise to stop the machine at any time during operation, press this button. An audible click
can be heard as the button locks in position.

The instrument should then be disconnected from the power supply.

Once disconnected, unlock the button by turning it clockwise. The instrument can then be
reconnected to the power supply and restarted.

3.5.4 Antifoam cartridge preparation

An empty 280mL volume antifoam cartridge is delivered as part of the installation kit. A recharge
bottle of antifoam is part of the package along with the necessary tools to transfer the liquid to
the cartridge (100mL volume syringe, tubing and connectors).

Note:

One antifoam recharge bottle should be enough to fill the cartridge at least 3 times and a cartridge

completely filled with antifoam is sufficient for approximately 3,000 measurements.

Important Note:

To optimize the performance and reliability of the instrument, it is strongly

recommended to only use the silicon antifoam supplied by Hach Lange (part number 33156).

To fill the cartridge, proceed as follows (numbers in the boxes correspond to the steps):

WARNING

Electrical danger and fire hazard. Only use the supplied power cable.

WARNING

Removable power cables must not be replaced with inadequately dimensioned power cables.

CAUTION

Check voltage requirement sticker on the rear panel of the instrument before connecting
(100-240 VAC).

Connect the power cable supplied with the instrument to the socket
at the rear of the instrument (left in diagram) for AC power
connection.

To switch the instrument on and off, press the rocker switch (right in
the diagram) - “I” for On and “O” for Off.