Maintenance

9.3 Maintenance menu overview

9.3.1 Wizards

This option gives an on-screen visual guide through a number of standard maintenance
procedures to perform on the instrument.

Installation

The installation wizard will be invoked automatically the first time the instrument is switched on to help
you complete the installation of the instrument.
Once the instrument has been successfully installed, the Skip installation wizard on device startup
checkbox should be checked to ensure a normal startup from there onwards.

Maintenance

It is recommended to use the maintenance wizards for maintenance of the instrument. Simply follow the
on-screen instructions available for each of the options.

Antifoam
cartridge
replacement

A replacement cartridge can be used or the existing cartridge refilled with antifoam. If
refilling the existing cartridge, follow the instructions given in

Antifoam cartridge

preparation on page 23

.

To ensure the performance and reliability of the instrument, it is strongly
recommended to only use the silicon antifoam supplied by Hach Lange (part number
33156).

EC sensor
maintenance

Chemical reactions within the gas sensor requires that the oxygen sensor be serviced
regularly to restore its original sensitivity. A clear sign that a sensor maintenance is
required is when measurements are noticeably less stable than usual.
For EC sensor maintenance, please also refer to the instruction sheet entitled A1100
cleaning procedure for 6110 TPA

which is supplied with the instrument.

Goretex filter
replacement

It is recommended to replace this filter every week.