Maintenance

9.3.2 Digital input verification

Piercing tip
replacement

It is recommended to replace these about every six months.
Note:

When replacing the tip, use the mounting tool supplied with the installation kit

and apply some Loctite

®

222 sealant (or similar) on the screw thread of the tip before

mounting.

Antifoam pump
maintenance

Refill or replace the antifoam cartridge as required.

Digital input verification

This allows you to check all devices that provide digital inputs to the instrument microprocessor. A black
dot in the circle to the right of the option signifies that the specific sensor has been activated.

Emergency stop pushed

Activated when the emergency stop button is pushed. The ultra
sound option will also have a black dot as the power supply of
the sonotrode is deactivated at the same time.

Axis reference position activated

The black dot indicates that the piercing device is positioned in
the reference (home) location.

End of travel upper bound activated

The upper bound limit is located higher than the reference
position. If this button is activated it means that the system is in
an abnormal state of operation and you should call your local
Hach Lange representative.

End of travel lower bound activated

Activated when the piercing device is in its lowest position.

Front door closed

Activated when the front door is closed.

Protection cover opened

Activated when the top protection cover is not in place.

Ultra sound error

Activated when the sonotrode is not working correctly.

Low level of antifoam

Activated when the antifoam cartridge is nearly empty.

Foam detection activated

Activated when foam or liquid has ingressed into the gas
sampling tube.

Bottle rim detection activated

Activated when a package is detected by the optical barrier. For
control purposes, this can be activated by passing the hand
between both optical detectors.