Section 6

Calibration

6.1 Calibration menu overview

This section describes the sequence of actions required to calibrate the sensors and flow meter
on the instrument:

1.

Barometric pressure sensor

2.

Pressure sensor

3.

Temperature sensor

4.

CO2 sensor

5.

Flow sensor

6.

O2 sensor

Note:

The above is the actual sequence of events that must be followed for any calibration (i.e. if you

calibrate the CO

2

sensor, then the barometric sensor, pressure sensor and temperature sensor must be

calibrated first and in that order).

6.2 Standards requirements

It is recommended that calibrations should be made with external certified standards at least
once a year.

The instrument will provide accurate results provided accurate reference standards are used.
The following table shows the maximum ranges for the standards. These ranges should not be
exceeded.

6.3 Calibration schedule

This following table shows the recommended sensor calibration intervals (using the instrument’s
automatic calibration feature) based on an average of 500 package analyses per week. This
proposed schedule can be modified according to operating conditions.

Standard

Range

Pressure

± 0.02 bar

Barometric pressure

± 2 mbar

Temperature

± 0.2°C

CO

2

sensor

CO

2

at 99.9% purity at ± 0.1% which is the purge gas for the normal operation

Sensor

Interval

Sensor

Interval

Barometric pressure sensor

6 months

O

2

sensor

2 months

Flow meter

6 months

CO

2

sensor

6 months

Temperature sensor

6 months

Pressure sensor

6 months