Measurement

7.1 Main screen

The initial screen displays standard information about the package to measure:

You have the option to change the details displayed or to start the measurement process:

User

Package ID

Start analysis

7.1.1 User

7.1.2 Package ID

7.1.3 Start analysis

Important Note:

Once a sample has completed the measurement process it will contain a small

amount of antifoam. Therefore, after measurement, ensure the sample is disposed of in a sink
(or similar) to avoid any risk of ingestion.

Note:

If the Diagnostic measurement view and the Display diagnostics results boxes have been checked

during configuration (see

Instrument parameters on page 37

), then the screens will differ (see examples in

Diagnostic measurement and results views on page 53

) but the process is the same.

For optimal operation, the base of the package holder should always have a layer of water for
improved transfer of the ultrasound energy to the liquid in the package. For the correct amount
of water, tilt the package holder forward by turning the tilt knob (No. 4 in

Figure 2 on page 15

) to

position 3 (small can icon). Next, fill the holder with water until the water level reaches the top
without spilling over. Finally, put the holder back to the correct tilt position for the package to be
measured in readiness for the measurement process.

User

Enter a valid User ID and Password. The new user name will be displayed.

Package ID

Define the package to be measured. The package details must have already been entered into the
system (see

Package parameters on page 35

for details).

Select the package from the list provided and press Ok. If necessary, you can change the package
details using the Edit button. For details on editing a package, refer to

Package parameters on page 35

).